If you overtighten the cap you may not be able to remove it, and you can damage your pump. You should not turn the cap more than four half turns.

  • Place the battery cap in the pump and tighten so the slot is aligned horizontally with the pump as shown here:Ĭaution: Do not overtighten the battery cap.
  • Allow cold batteries to reach room temperature before you insert them into your pump. Note: Do not use cold batteries, as the life of the battery may incorrectly appear as low. Check the label on the back of the pump to make sure the battery is inserted correctly. Put the new battery in the pump with the negative end going in first.
  • Remove the old battery and dispose of it per the disposable requirements of your state or country.
  • Turn the cap in a counter-clockwise direction.
  • Use the edge of a coin to remove the battery cap.
  • Do NOT remove the battery during a bolus or Fill Cannula delivery.
  • battery guardian mac

    Make sure the pump is at the HOME (idle) screen when you remove the battery.

    battery guardian mac

  • Clear ( ESC, ACT) any alarms and/or alerts before removing and replacing the battery.
  • After you clear the alarm, the pump will operate normally, with a decreased battery life.

    battery guardian mac

    The pump will suspend insulin delivery until you clear the alarm. If you receive a WEAK BATTERY alarm, respond to the alarm and continue.

    Battery guardian mac full#

    As a safety measure, if you insert a battery that does not have full power, the WEAK BATTERY or FAILED BATT TEST alarm may sound. Your pump accepts a new AAA alkaline battery, size E92, type LR03. Device: MiniMed ® 530G (551/751), MiniMed ® Paradigm Revel TM (523/723), MiniMed Paradigm ® 522/722, Guardian ® REAL-Time System, MiniMed Paradigm ® 515/715, MiniMed Paradigm ® 512/712









    Battery guardian mac